GO BACK

Ion Chromatography

ANALEST provides anion or cation identification via ion chromatography . Combining  background suppression (electrochemical regeneration),  temperature compensated conductivity detection and the Perkin Elmer Series 200 LC pump allows users to separate and quantify various ionic species in aqueous solutions. 

Ion Chromatography System:

PerkinElmer Series 200 binary pump

Alltech ERIS 1000HP Autosuppressor  (ERIS (Electrochemically Regenerated Ion Suppressor) 

Detection sensitivity is improved by both reducing the background conductivity of the eluent, while enhancing the signal of the analyte.  Using ERIS, the solid phase chemical suppressor is continually regenerated, allowing for reagent free operation.

Alltech 550 Conductivity Detector.

The conductivity detector has a temperature controlled cell compartment with a range of ambient to 60oC, which eliminates thermally induced baseline noise and drift.