Thursday - November 26, 2009Shimming issues on Bruker AvanceIII-400 solvedThe shimming issues on the Bruker Avance III 400 seem to be solved. Tuesday - November 17, 2009Dr Darcy BurnsDr Darcy Burns has joined the NMR Facility as Associate Manager. Monday - November 16, 2009Bruker BACS problems (updated)After just under 12,000 samples, the Bruker sample changer (BACS) is having problems. Thursday - November 05, 2009Bruker 400 (updated)The departmental Bruker AvanceIII-400 has some problems. First, Fluorine-19 tuning has become unreliable. Second, shimming also has become unreliable. Bruker personnel installed patches to tuning and shimming and these are working again. Thursday - November 05, 2009Staffing updates November 2009 (updated)Just a note on staffing. First, Dr Timothy Burrow has the flu and pneumonia and will not be back to work until Monday November 16th. Second, Ms Rose Balazs is on medical leave until mid-November. At this time only Dr Golembek is available and she will be in with her usual hours; 8:30-3 Monday, Tuesday and Thursday. Friday - October 30, 2009Varian VnmrS-400 spinning issuesThe inside of the magnet (what we call the upper barrel) was dirty and samples did not spin. This has been fixed. Friday - October 30, 2009Bruker issues on overnight samplesThere was an error running samples overnight last night, for samples submitted on October 29. Tuesday - October 27, 2009Bruker Avance IssuesTHere has been an intermittent error on the Bruker Avance III while running samples overnight. Many samples were not run. Last night, several samples were left in the queue, and not run. Friday - October 23, 2009Varian Mercury 400 issuesThe spinning rate control on the Varian Mercury 400 was not functioning correctly. A sticky valve was replaced and spinning is working again. Tuesday - September 22, 2009Varian Mercury 300 issuesThere have been some problems with the Mercury 300: On Friday, September 18, a brass hose connector came loose, causing incomplete ejection of samples and some broken NMR tubes. Today, we spent some time looking over the spectrometer. I found a loose solder connection on the proton pre-amplifier. This will be fixed in-house. Wednesday - September 16, 2009Mercury files not transferredThe server was patched and rebooted yesterday Sept 15 2009. For about 4 hours (10 AM to 2 PM) FIDs were not transferred. Wednesday - September 16, 2009Monday - September 14, 2009Helium fillsWe will be doing helium fills on the following spectrometers: Mercury 200, Mercury 300/SMS, Mercury 400/SMS, Unity 500 and Avance 400/BACS starting tomorrow, September 15 2009! Saturday - August 29, 2009Mercury 400 problemsCommunication between the console CPU and the Sun Sparcstation was lost last night ~ 9 PM. A faulty ethernet cable was replaced and the system is working again, 3 PM Saturday August 29. Monday - August 24, 2009Swap of AutoX probe on Varian VnmrS-400There is a problem with the tube switch selector on the probe in the Varian 400; the probe will be sent for repairs. We plan to swap the probe on Thursday August 27th, thus the spectrometer will be unavailable that day. Monday - August 24, 2009Thursday - August 20, 2009Monday - August 17, 2009Mercury 400 spinning issues [solved]There is a problem with the spinner control on the Varian Mercury 400. The part is is only about 25 years old... Thursday - August 13, 2009Bruker AVIII-400 with BACS working againAccording to the schedule, the Bruker AVIII-400 with BACS was checked. Also, not according to schedule, a leaking condensation pump in a room across the hall flooded the floor around the system with dirty water. Tuesday - August 11, 2009Mercury 400 issues resolvedThis morning, a tube that was cracked on the bottom was found on the sample changer rack of the Mercury 400 this morning. Because it was unclear if the crack was there before or after the sample was run, as a precautionary measure, the probe and upper barrel were washed and cleaned. |