Thursday - November 26, 2009

 Shimming issues on Bruker AvanceIII-400 solved


 The shimming issues on the Bruker Avance III 400 seem to be solved.


Posted at 10:10 AM     Read More  

Tuesday - November 17, 2009

 Dr Darcy Burns


 Dr Darcy Burns has joined the NMR Facility as Associate Manager. 


Posted at 05:22 PM     Read More  

Monday - November 16, 2009

Bruker BACS problems (updated)


 After just under 12,000 samples, the Bruker sample changer (BACS) is having problems.


Posted at 06:13 PM     Read More  

Thursday - November 05, 2009

 Bruker 400 (updated)


 The departmental Bruker AvanceIII-400 has some problems. First, Fluorine-19 tuning has become unreliable. Second, shimming also has become unreliable.


Bruker personnel installed patches to tuning and shimming and these are working again.


Posted at 11:34 PM     Read More  

Thursday - November 05, 2009

 Staffing updates November 2009 (updated)


 Just a note on staffing. First, Dr Timothy Burrow has the flu and pneumonia and will not be  back to work until Monday November 16th. Second, Ms Rose Balazs is on medical leave until mid-November. At this time only Dr Golembek is available and she will be in with her usual hours; 8:30-3 Monday, Tuesday and Thursday.


Posted at 11:31 PM     Read More  

Friday - October 30, 2009

 Varian VnmrS-400 spinning issues


The inside of the magnet (what we call the upper barrel) was dirty and samples did not spin. This has been fixed.


Posted at 11:47 AM     Read More  

Friday - October 30, 2009

 Bruker issues on overnight samples


 There was an error running samples overnight last night, for samples submitted on October 29.


Posted at 11:45 AM     Read More  

Tuesday - October 27, 2009

 Bruker Avance Issues


 THere has been an intermittent error on the Bruker Avance III while running samples overnight. Many samples were not run. Last night, several samples were left in the queue, and not run.


Posted at 11:22 AM     Read More  

Friday - October 23, 2009

 Varian Mercury 400 issues


 The spinning rate control on the Varian Mercury 400 was not functioning correctly. A sticky valve was replaced and spinning is working again.


Posted at 10:41 AM     Read More  

Tuesday - September 22, 2009

 Varian Mercury 300 issues


 There have been some problems with the Mercury 300:

On Friday, September 18, a brass hose connector came loose, causing incomplete ejection of samples and some broken NMR tubes.

Today, we spent some time looking over the spectrometer. I found a loose solder connection on the proton pre-amplifier. This will be fixed in-house.


Posted at 05:53 PM     Read More  

Wednesday - September 16, 2009

 Mercury files not transferred


 The server was patched and rebooted yesterday Sept 15 2009. For about 4 hours (10 AM to 2 PM) FIDs were not transferred.


Posted at 03:49 PM     Read More  

Wednesday - September 16, 2009

 All helium fills done


 All helium fills done.


Posted at 03:49 PM     Read More  

Monday - September 14, 2009

 Helium fills


 We will be doing helium fills on the following spectrometers: Mercury 200, Mercury 300/SMS, Mercury 400/SMS, Unity 500 and Avance 400/BACS starting tomorrow, September 15 2009!


Posted at 03:32 PM     Read More  

Saturday - August 29, 2009

 Mercury 400 problems


Communication between the console CPU and the Sun Sparcstation was lost last night ~ 9 PM. 

A faulty ethernet cable was replaced and the system is working again, 3 PM Saturday August 29.


Posted at 03:01 PM     Read More  

Monday - August 24, 2009

Swap of AutoX probe on Varian VnmrS-400


There is a problem with the tube switch selector on the probe in the Varian 400; the probe will be sent for repairs. We plan to swap the probe on Thursday August 27th, thus the spectrometer will be unavailable that day.


Posted at 09:49 AM     Read More  

Monday - August 24, 2009

 Rose Balazs


 Ms Rose Balazs will be away until September 10th.


Posted at 09:47 AM     Read More  

Thursday - August 20, 2009

 NMR tube policy


 A brief summary of the NMR tube policy.


Posted at 08:59 AM     Read More  

Monday - August 17, 2009

 Mercury 400 spinning issues [solved]


 There is a problem with the spinner control on the Varian Mercury 400. The part is is only about 25 years old...


Posted at 12:01 PM     Read More  

Thursday - August 13, 2009

Bruker AVIII-400 with BACS working again


 According to the schedule, the Bruker AVIII-400 with BACS was checked.

Also, not according to schedule, a leaking condensation pump in a  room across the hall flooded the floor around the system with dirty water.


Posted at 04:50 PM     Read More  

Tuesday - August 11, 2009

 Mercury 400 issues resolved


This morning, a tube that was cracked on the bottom was found on the sample changer rack of the Mercury 400 this morning. Because it was unclear if the crack was there before or after the sample was run, as a precautionary measure, the probe and upper barrel were washed and cleaned.


Posted at 05:35 PM     Read More  


©