Monday - January 11, 2010

 Software on the Varian Mercury 200 reloaded


 After January 4th, VnmrJ on the mercury 200 was not starting up correctly. The software was re-installed and this is fixed.


Posted at 11:07 AM     Read More

Monday - January 11, 2010

 MestReNova printing


 Printing from MestReNova is very slow on "keroppi".


Posted at 11:04 AM     Read More

Monday - December 07, 2009

 NMR Status page


 The NMR Facility is pleased to announce the new NMR Status page. It shows the status of all the NMR spectrometers and the state of the queues on the sample changers.


Posted at 09:46 AM     Read More

Thursday - November 26, 2009

 Shimming issues on Bruker AvanceIII-400 solved


 The shimming issues on the Bruker Avance III 400 seem to be solved.


Posted at 10:10 AM     Read More

Monday - November 23, 2009

 MestreNova 5 Lite is available with ChemOffice 12


 All users in the University of Toronto can download a free yearly license for ChemOffice 12 from the Chemistry Library. This also includes a copy of MestreNova 5 Lite for both Windows and Macintosh. Go to the Chemistry Library for more information.


Posted at 09:45 AM     Read More

Tuesday - November 17, 2009

 Dr Darcy Burns


 Dr Darcy Burns has joined the NMR Facility as Associate Manager. 


Posted at 05:22 PM     Read More

Monday - November 16, 2009

Bruker BACS problems (updated)


 After just under 12,000 samples, the Bruker sample changer (BACS) is having problems.


Posted at 06:13 PM     Read More

Thursday - November 05, 2009

 Bruker 400 (updated)


 The departmental Bruker AvanceIII-400 has some problems. First, Fluorine-19 tuning has become unreliable. Second, shimming also has become unreliable.


Bruker personnel installed patches to tuning and shimming and these are working again.


Posted at 11:34 PM     Read More

Thursday - November 05, 2009

 Staffing updates November 2009 (updated)


 Just a note on staffing. First, Dr Timothy Burrow has the flu and pneumonia and will not be  back to work until Monday November 16th. Second, Ms Rose Balazs is on medical leave until mid-November. At this time only Dr Golembek is available and she will be in with her usual hours; 8:30-3 Monday, Tuesday and Thursday.


Posted at 11:31 PM     Read More

Friday - October 30, 2009

 Tuning


Some users are confused over tuning and what it does. Some probes appear not to need tuning whereas the newer systems spend some time tuning.


Posted at 01:23 PM     Read More

Friday - October 30, 2009

 Varian VnmrS-400 spinning issues


The inside of the magnet (what we call the upper barrel) was dirty and samples did not spin. This has been fixed.


Posted at 11:47 AM     Read More

Friday - October 30, 2009

 Bruker issues on overnight samples


 There was an error running samples overnight last night, for samples submitted on October 29.


Posted at 11:45 AM     Read More

Wednesday - October 28, 2009

 MestreNova on Keroppi upgraded


 I've upgraded the copy of MestreNova on Keroppi to version 6.0.2, the latest version. Version 6 has some nice additions that would be useful for all chemists, see below (Images stolen from MestreNova LLC):


Posted at 09:36 AM     Read More

Tuesday - October 27, 2009

 Bruker Avance Issues


 THere has been an intermittent error on the Bruker Avance III while running samples overnight. Many samples were not run. Last night, several samples were left in the queue, and not run.


Posted at 11:22 AM     Read More

Friday - October 23, 2009

 Varian Mercury 400 issues


 The spinning rate control on the Varian Mercury 400 was not functioning correctly. A sticky valve was replaced and spinning is working again.


Posted at 10:41 AM     Read More

Tuesday - October 06, 2009

 New associate manager


 We are please to say that an associate manager was hired and will start in November.


Posted at 07:28 PM     Read More

Tuesday - October 06, 2009

 Manager away until October 19th


 Dr Timothy Burrow will be away until October 19th. Dr Adina Golombek will be in as usual, on Mondays, Tuesdays and Thursday until 3PM. Ms Rose Balazs is away until October 12. Dmitry will reset the sample queues in the morning. 


Posted at 07:25 PM     Read More

Tuesday - September 22, 2009

 Varian Mercury 300 issues


 There have been some problems with the Mercury 300:

On Friday, September 18, a brass hose connector came loose, causing incomplete ejection of samples and some broken NMR tubes.

Today, we spent some time looking over the spectrometer. I found a loose solder connection on the proton pre-amplifier. This will be fixed in-house.


Posted at 05:53 PM     Read More

Wednesday - September 16, 2009

 Mercury files not transferred


 The server was patched and rebooted yesterday Sept 15 2009. For about 4 hours (10 AM to 2 PM) FIDs were not transferred.


Posted at 03:49 PM     Read More

Wednesday - September 16, 2009

 All helium fills done


 All helium fills done.


Posted at 03:49 PM     Read More

Monday - September 14, 2009

 Helium fills


 We will be doing helium fills on the following spectrometers: Mercury 200, Mercury 300/SMS, Mercury 400/SMS, Unity 500 and Avance 400/BACS starting tomorrow, September 15 2009!


Posted at 03:32 PM     Read More

Friday - September 11, 2009

 Bruker Avance III-400 P31 shift error


 All Phosphorus-31 chemical shifts of samples run in C6D6 have an error! The chemical shifts are shifted downfield by 13.3560 ppm.


Posted at 10:11 AM     Read More

Wednesday - September 09, 2009

 More server fall-out


Some problems with moving to backup hardware after the server failed. 

FTP was not working

Printing to "babe" was not working

Mail out of NMR was not working until this morning.


Update September 10/2009 11 AM.

The departmental proxy "grandial" was being blocked. Only windows users within the department were affected and could not access the booking schedule and training signup. This has been fixed. 

Files from the Mercury 300 and 400 were not being copied for ~24h, an errant space in the script was removed and files are being copied again.


Update September 10/2009 1 PM.

Bruker files now transferring!



Posted at 03:54 PM     Read More

Tuesday - September 08, 2009

 Server problems


Because the Data drive (where all user data is stored) was reaching capacity, I copied all data to a new RAID system and installed in the server. All the data is OK, but the system did not boot. 

[6:15 PM most systems are now back online. The server had major hardware problems]

[7:15 PM all spectrometers working, all data stations working]


Posted at 10:11 AM     Read More

Tuesday - September 01, 2009

 Plotting colour issues


 There are problems with colour plotting from Keroppi. Follow these instructions to get the colours you want.


Posted at 05:18 PM     Read More

Monday - August 31, 2009

 Keroppi finally fixed


 The workstation Keroppi is working again. It has a copy of MNova and VnmrJ for users to process NMR data.


Posted at 03:12 PM     Read More

Monday - August 31, 2009

 NMR server problems


 The boot hard drive on the NMR server had problems starting ~9PM Saturday August 29. The drive was fixed by 11:45 PM.


Posted at 11:09 AM     Read More

Saturday - August 29, 2009

 Mercury 400 problems


Communication between the console CPU and the Sun Sparcstation was lost last night ~ 9 PM. 

A faulty ethernet cable was replaced and the system is working again, 3 PM Saturday August 29.


Posted at 03:01 PM     Read More

Tuesday - August 25, 2009

 Connection to "nmrweb" blocked


 People within chemistry cannot connect to the nMR booking server, https://nmrweb.chem.utoronto.ca/ This is because of a configuration of the departmental squid proxy (grandial). 


Posted at 09:54 AM     Read More

Tuesday - August 25, 2009

 Mercury 400 problems


 The old software/sun computer system have had some problems lately. 


Posted at 08:59 AM     Read More

Monday - August 24, 2009

Swap of AutoX probe on Varian VnmrS-400


There is a problem with the tube switch selector on the probe in the Varian 400; the probe will be sent for repairs. We plan to swap the probe on Thursday August 27th, thus the spectrometer will be unavailable that day.


Posted at 09:49 AM     Read More

Monday - August 24, 2009

 Rose Balazs


 Ms Rose Balazs will be away until September 10th.


Posted at 09:47 AM     Read More

Thursday - August 20, 2009

 NMR tube policy


 A brief summary of the NMR tube policy.


Posted at 08:59 AM     Read More

Monday - August 17, 2009

 Mercury 400 spinning issues [solved]


 There is a problem with the spinner control on the Varian Mercury 400. The part is is only about 25 years old...


Posted at 12:01 PM     Read More

Thursday - August 13, 2009

Bruker AVIII-400 with BACS working again


 According to the schedule, the Bruker AVIII-400 with BACS was checked.

Also, not according to schedule, a leaking condensation pump in a  room across the hall flooded the floor around the system with dirty water.


Posted at 04:50 PM     Read More

Wednesday - August 12, 2009

 Workstation Keroppi problems


 The workstation keroppi suffered a hard drive failure. 


Posted at 05:42 PM     Read More

Wednesday - August 12, 2009

 NMR server blocked


 After a misconfiguration, the NMR server timothy was blocked by Frank Bures. We are waiting for it to be unblocked.

[AT 4 PM, the server was unblocked]


Posted at 11:03 AM     Read More

Tuesday - August 11, 2009

 Mercury 400 issues resolved


This morning, a tube that was cracked on the bottom was found on the sample changer rack of the Mercury 400 this morning. Because it was unclear if the crack was there before or after the sample was run, as a precautionary measure, the probe and upper barrel were washed and cleaned.


Posted at 05:35 PM     Read More

Tuesday - August 11, 2009

 Server updates


 The main NMR server will be updated today for security patches, with reboot around 10 AM.

[Update: server rebooted and all services restored.]


Posted at 09:44 AM     Read More

Thursday - August 06, 2009

 Deuterium NMR


 Deuterium NMR has been added to the Varian VnmrS-400. 


Posted at 03:58 PM     Read More

Thursday - August 06, 2009

 Carbon with Fluorine decoupling


 Carbon spectra with fluorine decoupling has been added to the VnmrS-400. This used garp decoupling to decouple from -100 to -175 ppm (the usual organic range). To change the decoupling window, set the "dof" parameter appropriately.

Click on the CarbonDECFluorine button on the experiment tab. There is also Phosphorus and Boron with Fluorine decoupling, which are PhosphorusDECFluorine and BoronDECFluorine, respectively.

Note that since the probe is only two channel, there is no proton decoupling; we need a H/F/X triple probe to decouple both proton and fluorine.


Posted at 03:49 PM     Read More

Tuesday - July 21, 2009

 Work on Mercury 300


There will be no access to the Mercury 300 on July 29 and July 30 for routine maintenance and computer upgrade.


Posted at 02:51 PM     Read More

Tuesday - July 21, 2009

 Helium Fills July 27, 2009


 Access to the Mercury 200, Unity 400 and Unity 500 will be limited on Monday July 27, 2009 for LHe fills.


Posted at 02:37 PM     Read More

Wednesday - June 24, 2009

 Helium Fills Thursday June 25


Due to Helium fills, access to the Mercury 300 and Mercury 400 will limited in the afternoon 2 PM onwards. Access to the Varian 400 will be limited on Friday morning.


Posted at 09:25 AM     Read More

Tuesday - June 23, 2009

 Centre for Spectroscopic Investigation of Complex Organic Molecules and Polymers (CSICOMP) funded!


 The CFI grant forwarded by Mark Lautens has been funded by CFI and the University's Infrastructure Operating Fund for a total of $3,365,393.


Posted at 09:47 AM     Read More

Friday - June 12, 2009

 Varian VnmrS-400 available again


 The Varian VnmrS-400 is available again.


Posted at 07:31 PM     Read More

Wednesday - June 10, 2009

 NMR Server rebooted


 From 5 PM to 6PM today (June 10 2009) the server was rebooted for patches.


Posted at 06:04 PM     Read More

Wednesday - June 10, 2009

 Files transferred from the Bruker


 Some files were transferring from the Bruker while others were not.


Posted at 04:29 PM     Read More

Tuesday - June 09, 2009

 Varian 400


 We have replaced the faulty part (frequency generator) on the Varian 400.


Posted at 06:59 PM     Read More

Tuesday - June 09, 2009

 NMR calendar


 The NMR calendar of events


Posted at 06:57 PM     Read More


©